DocumentCode
789876
Title
Test of Six High Temperature Neutron Detectors to 650°C
Author
Bacastow, J.L. ; Swickard, E.O.
Author_Institution
University of California Los Alamos Scientific Laboratory Los Alamos, New Mexico
Volume
16
Issue
1
fYear
1969
Firstpage
177
Lastpage
183
Abstract
Two neutron detectors (fission counters) from each of three manufacturers were tested at several temperatures between ambient and 650° C.. Parameters examined as a function of temperature were: leakage current, capacitance, noise vs. voltage, and pulse height distribution.
Keywords
Capacitance; Counting circuits; Detectors; Leak detection; Leakage current; Manufacturing; Neutrons; Temperature; Testing; Voltage;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1969.4325103
Filename
4325103