DocumentCode
789913
Title
Performance Tests on an Eddy-Current Flowmeter
Author
Wiegand, David E. ; Michels, Charles W.
Author_Institution
Magnetics Group, IEEE Argonne National Laboratory, 9700 S. Cass Ave., Argonne, Illinois, 60439
Volume
16
Issue
1
fYear
1969
Firstpage
192
Lastpage
195
Abstract
The eddy-current flowmeter seems well adapted to measurements in high-temperature liquid-metal systems, where deterioration of the properties of permanent magnets and other ferromagnetic parts or bearing problems may seriously degrade the performance of other types. Tests on an eddy-current flowmeter confirm the high signal levels and the optimum excitation frequency predicted by an analysis of the device.
Keywords
Aluminum; Coils; Distortion measurement; Fluid flow measurement; Frequency; Laboratories; Magnetic analysis; Testing; Velocity measurement; Voltage;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1969.4325106
Filename
4325106