DocumentCode
790225
Title
A Low Level Environmental Radiation Monitor with High Sensitivity, Stability, and Dynamic Range
Author
Cassada, Thomas E.
Author_Institution
Electronics Section Southwestern Radiological Health Laboratory
Volume
16
Issue
1
fYear
1969
Firstpage
405
Lastpage
410
Abstract
An environmental radiation monitor is described which records gamma radiation levels with a logarithmic response over a dynamic range of four decades (.01 mR/hr to 100 mR/hr achievable without range switching). Constant temperature accuracy is greater than ±10% of indicated reading throughout 3.5 decades of the four decade range. A high degree of temperature stability is achieved (error due to temperature change is less than ±20% of reading over an ambient temperature range of -10°C to + 55°C). Solid-state electronic circuitry is utilized exclusively for increased reliability and long term stability.
Keywords
Detectors; Dynamic range; Filling; Gamma rays; Mercury (metals); Radiation monitoring; Solid state circuits; Stability; Temperature distribution; Voltage;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1969.4325136
Filename
4325136