  • Abstract
    In this paper, we present a radiotelemetry sensor, designed as a lab-in-a-pill, which incorporates a two-channel microfabricated sensor platform for real-time measurements of temperature and pH. These two parameters have potential application for use in remote biological sensing (for example they may be used as markers that reflect the physiological environment or as indicators for disease, within the gastrointestinal tract). We have investigated the effects of biofouling on these sensors, by exploring their response time and sensitivity in a model in vitro gastrointestinal system. The artificial gastric and intestinal solutions used represent a model both for fasting, as well as for the ingestion of food and subsequent digestion to gastrointestinal chyme. The results showed a decrease in pH sensitivity after exposure of the sensors for 3 h. The response time also increased from an initial measurement time of 10 s in pure GI juice, to ca. 25 s following the ingestion of food and 80 s in simulated chyme. These in vitro results indicate that changes in viscosity in our model gastrointestinal system had a pronounced effect on the unmodified sensor
  • Keywords
    chemical sensors; diseases; microsensors; patient diagnosis; radiotelemetry; temperature sensors; viscosity; artificial gastrointestinal environments; biocompatibility; biofouling; disease; gastrointestinal system; gastrointestinal tract; lab-on-a-pill sensor; pH measurement; radiotelemetry sensor; remote biological sensing; temperature measurement; two-channel microfabricated sensor platform; viscosity; Biological system modeling; Biosensors; Delay; Diseases; Gastrointestinal tract; In vitro; Intestines; Sensor systems; Temperature measurement; Temperature sensors; Biofouling; in vitro measurements; lab-in-a-pill; microfabricated sensors; radio telemetry; Biocompatible Materials; Biomimetic Materials; Equipment Design; Equipment Failure Analysis; Gastrointestinal Contents; Gastrointestinal Tract; Hydrogen-Ion Concentration; Ion-Selective Electrodes; Prostheses and Implants; Reproducibility of Results; Sensitivity and Specificity; Telemetry; Thermography; Transducers;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2006.883698
  • Filename
    1710175