• DocumentCode
    794610
  • Title

    Evaluation of Corneal Refraction in a Model of a Gaze Tracking System

  • Author

    Villanueva, Arantxa ; Cabeza, Rafael

  • Author_Institution
    Dept. of Electr. & Electron. Eng., Public Univ. of Navarra, Pamplona
  • Volume
    55
  • Issue
    12
  • fYear
    2008
  • Firstpage
    2812
  • Lastpage
    2822
  • Abstract
    The complexity of corneal refraction modeling complicates geometrical analysis of gaze tracking systems. The task can be accomplished in different ways by means of different approximations. Depending on the assumptions made, errors can arise and the accuracy of trackers can end up determined to an excessive degree by the hardware setup and subject parameters. In this study, keeping the hardware and eye model the same, we compare different methods for evaluation of corneal refraction. Our results show that the approximated methods are sensitive to the calibration of the system and less robust to noise. The nonapproximated method removes any dependency on setup and allows clear differentiation between hardware and subject´s calibration.
  • Keywords
    biocontrol; eye; geometrical optics; light refraction; man-machine systems; corneal refraction evaluation; corneal refraction modeling; gaze tracking system; geometrical optics; Calibration; Cameras; Geometry; Hardware; Human computer interaction; Kinematics; Noise robustness; Optical refraction; Physiology; Reflection; Solid modeling; Corneal refraction; gaze tracking; Artifacts; Calibration; Cornea; Corneal Topography; Eye Movement Measurements; Eye Movements; Humans; Image Interpretation, Computer-Assisted; Models, Biological; Refraction, Ocular; Retinoscopy;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2008.2002152
  • Filename
    4564194