A Digital Closed-Loop Calibration System for Space-Borne Particle Spectrometers
Author :
Streeter, Gary R. ; Guenther, G.A.
Author_Institution :
Space and Tactical Systems Corporation One Garfield Circle, Burlington, Mass. 01803
Volume :
17
Issue :
1
fYear :
1970
Firstpage :
124
Lastpage :
129
Abstract :
The operation of a closed-loop, in-flight calibration system is described. This system utilizes a Cesium Iodide crystal with a grown-in Am241 source as the calibration reference and a programmable photomultiplier dynode supply as the gain-change element. Through the use of this system, the overall gain of a particle spectrometer is expected to be kept accurate to within 3% throughout the changing orbital environment. Approximate power consumption of the system, including the dynode supply, analog section and digital sectoring and control is 415 milliwatts.
Keywords :
Calibration; Circuits; Detectors; Energy consumption; Instruments; Photomultipliers; Plastics; Spectroscopy; Telescopes; Voltage;