DocumentCode :
794900
Title :
Low-Current Measurements Using Transistor Logarithmic DC Electrometers
Author :
Kennedy, E.J.
Author_Institution :
Oak Ridge National Laboratory Oak Ridge, Tennessee 37830
Volume :
17
Issue :
1
fYear :
1970
Firstpage :
326
Lastpage :
334
Abstract :
A nine-decade (10-13 to 10-4 amp) and a five-decade (4 ?? 10-12 to 4 ?? 10-7 amp) dc logarithmic electrometer have been designed, and their performance evaluated. The problems of drift, linear-to-log conversion accuracy, temperature dependence, and frequency response are considered. An accuracy of better than ??0.6%/decade (~2% of reading) was achieved.
Keywords :
Bipolar transistors; Circuits; Feedback; Frequency response; Laboratories; Semiconductor diodes; Silicon compounds; Temperature dependence; Temperature sensors; Voltage;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1970.4325594
Filename :
4325594
Link To Document :
بازگشت