DocumentCode
795884
Title
Spectrum Degradation Effects in Ge(Li) Detectors
Author
Armantrout, G.A. ; Thompson, R.W., Jr.
Author_Institution
University of California, Lawrence Radiation Laboratory Livermore, California 94550
Volume
17
Issue
3
fYear
1970
fDate
6/1/1970 12:00:00 AM
Firstpage
165
Lastpage
175
Abstract
The mechanisms responsible for the degradation of the performance of Ge(Li) detectors (which include the effects of uniform trapping, non-uniform trap distribution, and locally weak collection fields) are considered, and simulated spectra have been computed. Experimental measurements have been made to identify these sources of degradation. The important results include a measurement of the effective capture cross sections of the most important traps and a determination of the main sources of non-uniform trap distribution and weak collection fields in Ge(Li) detectors.
Keywords
Crystals; Degradation; Energy states; Infrared detectors; Laboratories; Lithium; Measurement techniques; Photonic band gap; Radiation detectors; Temperature;