• DocumentCode
    797046
  • Title

    Removal of power line interference in signal-averaged electrocardiography systems

  • Author

    Ider, Y. Ziya ; Saki, M. Cem ; Gucer, H. Alper

  • Author_Institution
    Dept. of Elctr. Eng., Middle East Tech. Univ., Ankara, Turkey
  • Volume
    42
  • Issue
    7
  • fYear
    1995
  • fDate
    7/1/1995 12:00:00 AM
  • Firstpage
    731
  • Lastpage
    735
  • Abstract
    A method for line interference reduction to be used in signal-averaged electrocardiography (SAECG) systems is proposed and its performance is analyzed. This new method is an adaptation of a previously reported technique for removal of line interference from conventional electrocardiograms. It involves the recording of a line reference signal simultaneous with the lead signals, so that a shifted and sealed version of it can be used to subtract line interference from the leads. It is shown that this line interference subtraction method can reduce line interference effectively and without introducing any additional noise into the ECG signal. It is also shown that Late Potential diagnostic decisions are not altered when this filter is applied. It is recommended that this technique be used in SAECG when line interference is unavoidable.
  • Keywords
    electrocardiography; interference (signal); medical signal processing; ECG signal noise; filter; late potential diagnostic decisions; lead signals; line interference reduction; line reference signal; power line interference removal; signal-averaged electrocardiography systems; Calibration; Chemistry; Diabetes; Electrocardiography; Geometry; Interference; Noise level; Noise reduction; Raman scattering; Spectroscopy; Electrocardiography; Filtration; Humans; Signal Processing, Computer-Assisted;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.391173
  • Filename
    391173