DocumentCode
797702
Title
A Laboratory Quality Charge Sensitive Amplifier Suitable for Space Applications
Author
Alexander, R.M.
Author_Institution
TRW Systems Redondo Beach, California
Volume
18
Issue
1
fYear
1971
Firstpage
264
Lastpage
274
Abstract
The amplifier described is designed for use with uncooled silicon nuclear particle detectors connonly used in space instruments. The charge sensitive preamplifier and two-stage shaping amplifier provides the performance of a laboratory quality instrument while simultaneously satisfying the stringent requirements imposed upon flight hardware. Double differentiation-double integration (DDDI) pulse shaping with pole-zero cancellation of the preamplifier decay time constant was selected on the basis of simplicity, adequate noise performance, and stabillity at high count rates. The components used in the design are compatible with hybrid microcircuit fabrication. Noise performance at 1 ??sec DDDI time constants was measured to be 1.8 KeV + 47 eV/pf FWHM (Silicon) for an amplifier consuming 100 mW. The operating temperature range is approximately -40?? to +80??C. Further results of an extensive testing program are reported.
Keywords
Hardware; Instruments; Laboratories; Noise cancellation; Noise shaping; Preamplifiers; Pulse amplifiers; Radiation detectors; Silicon; Space charge;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1971.4325875
Filename
4325875
Link To Document