• DocumentCode
    79899
  • Title

    Neutron Spectrometry with EJ299-33 Plastic Scintillator for {E_n} = 10 ! - ! 100 ~\\hbox {MeV}

  • Author

    Buffler, A. ; Comrie, A.C. ; Smit, F.D. ; Wortche, H.J.

  • Author_Institution
    Dept. of Phys., Univ. of Cape Town, Rondebosch, South Africa
  • Volume
    62
  • Issue
    3
  • fYear
    2015
  • fDate
    Jun-15
  • Firstpage
    1422
  • Lastpage
    1428
  • Abstract
    A neutron spectrometer based on a single EJ299-33 plastic scintillator (5 cm diameter × 5 cm) is described. A digital implementation of pulse shape discrimination is used to separate events associated with neutrons from those associated with gamma-rays. Measurements made using a ns-pulsed neutron beam are used to produce response functions of the detector for neutrons over the energy range 10 - 100 MeV. These lineshapes are used to test the capability of the spectrometer to produce neutron energy spectra, via the unfolding of pulse height spectra, for use in neutron fields having any type of time structure. The results are compared with those from a similar spectrometer constructed from EJ301 liquid scintillator.
  • Keywords
    liquid scintillation detectors; neutron spectrometers; solid scintillation detectors; EJ299-33 plastic scintillator; EJ301 liquid scintillator; neutron energy spectra; neutron fields; neutron spectrometry; ns-pulsed neutron beam; pulse shape discrimination; time structure; Detectors; Energy measurement; Gamma-rays; Neutrons; Particle beams; Plastics; Shape; Neutron spectrometry; pulse shape discrimination; scintillation detectors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2015.2430455
  • Filename