• DocumentCode
    80104
  • Title

    Innovative design of implanted medical lead to reduce MRI-induced scattered electric fields

  • Author

    Das, Ratan ; Yoo, Hyoungsuk

  • Author_Institution
    Dept. of Biomed. Eng., Univ. of Ulsan, Ulsan, South Korea
  • Volume
    49
  • Issue
    5
  • fYear
    2013
  • fDate
    February 28 2013
  • Firstpage
    323
  • Lastpage
    324
  • Abstract
    The radio frequency coils in magnetic resonance imaging (MRI) systems induce scattered electric fields in the implanted medical lead and result in tissue heating. A novel structure of the medical lead has been designed to reduce the scattered field. Proposed metal nails are placed along the medical lead. The scattered electric field is calculated near the medical lead tip at 64 MHz (1.5 T MRI). By installing this design on the conventional lead, a significant decrease in the scattered electric field at the tip of the lead is observed.
  • Keywords
    biological tissues; biomedical MRI; coils; MRI induced scattered electric field; frequency 64 MHz; implanted medical lead; magnetic flux density 1.5 T; magnetic resonance imaging; radio frequency coil; tissue heating;
  • fLanguage
    English
  • Journal_Title
    Electronics Letters
  • Publisher
    iet
  • ISSN
    0013-5194
  • Type

    jour

  • DOI
    10.1049/el.2012.4033
  • Filename
    6473935