DocumentCode
80104
Title
Innovative design of implanted medical lead to reduce MRI-induced scattered electric fields
Author
Das, Ratan ; Yoo, Hyoungsuk
Author_Institution
Dept. of Biomed. Eng., Univ. of Ulsan, Ulsan, South Korea
Volume
49
Issue
5
fYear
2013
fDate
February 28 2013
Firstpage
323
Lastpage
324
Abstract
The radio frequency coils in magnetic resonance imaging (MRI) systems induce scattered electric fields in the implanted medical lead and result in tissue heating. A novel structure of the medical lead has been designed to reduce the scattered field. Proposed metal nails are placed along the medical lead. The scattered electric field is calculated near the medical lead tip at 64 MHz (1.5 T MRI). By installing this design on the conventional lead, a significant decrease in the scattered electric field at the tip of the lead is observed.
Keywords
biological tissues; biomedical MRI; coils; MRI induced scattered electric field; frequency 64 MHz; implanted medical lead; magnetic flux density 1.5 T; magnetic resonance imaging; radio frequency coil; tissue heating;