DocumentCode
803799
Title
Evaluation of clinical PET count rate performance
Author
Watson, Charles C. ; Casey, Michael E. ; Beyer, Thomas ; Bruckbauer, Thomas ; Townsend, David W. ; Brasse, David
Author_Institution
CPS Innovations, Knoxville, TN, USA
Volume
50
Issue
5
fYear
2003
Firstpage
1379
Lastpage
1385
Abstract
We propose a new methodology for analyzing clinical positron emission tomography (PET) count rate performance that involves matching net trues (prompts - delayeds) and randoms (delayeds) count rate responses measured on a reference phantom to the actual patient data. We employ these estimated response curves to compute a performance metric closely related to the noise equivalent count rate, as a function of the total singles event rate in the system. From this we can determine the peak performance value relative to the measured performance for any individual acquisition. This maximum performance value depends only on the emitter and attenuator distributions. These peak count rates can be used to derive frame durations for equivalent noise, and correlated with patient weight. We can also determine the singles rate and, thereby, activity necessary to achieve this maximum performance and, based on the known activity in the patient, predict optimal injected doses.
Keywords
dosimetry; positron emission tomography; clinical PET count rate performance; estimated response curves; frame durations; noise equivalent count rate; optimal injected doses; positron emission tomography; Anthropomorphism; Attenuators; Biomedical imaging; Delay; Imaging phantoms; Measurement standards; Medical diagnostic imaging; Performance analysis; Positron emission tomography; Scattering;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.2003.817314
Filename
1236936
Link To Document