Abstract :
We describe a camera for the laminographic imaging of the thyroid without administration of radiopharmaceuticals to the patient. An external source of gamma-rays is used to excite the characteristic x-rays of natural iodine in the patient´s thyroid, source geometry limiting excitation to well-defined planes. The camera consists of a parallel hole collimator and a xenon-filled proportional wire chamber with digitized readout of coordinates. Pulse height selection is provided to limit events in the image display to a selected energy range. We estimate that the system will obtain high resolution laminography for local exposures on the order of 1 to 5 rad, with exposure times of a few minutes for each laminogram.
Keywords :
Cameras; Collimators; Energy resolution; High-resolution imaging; Instruments; Laboratories; Optical imaging; Wire; X-ray detection; X-ray detectors;