• DocumentCode
    805769
  • Title

    Electric field of a six-needle array electrode used in drug and DNA delivery in vivo: analytical versus numerical solution

  • Author

    Dev, Sukhendu B. ; Dhar, Deepak ; Krassowska, Wanda

  • Author_Institution
    Gene Delivery & Expression Sci., San Diego, CA, USA
  • Volume
    50
  • Issue
    11
  • fYear
    2003
  • Firstpage
    1296
  • Lastpage
    1300
  • Abstract
    We present an analytical solution for the electrical potential and field established by a six-needle array electroporation electrode, which is used in vivo for cancer treatment and DNA delivery. The analytical solution closely matches the numerical solution obtained with the finite element method: the mean error is less than 0.6 %.
  • Keywords
    DNA; biomedical electrodes; cancer; drug delivery systems; electric fields; finite element analysis; patient treatment; DNA delivery; analytical solution; cancer treatment; electroporation therapy; finite element method; mean error; numerical solution; pulsed electric fields; six-needle array electrode; Biomedical electrodes; Cancer; DNA; Drugs; Electric potential; Finite element methods; In vivo; Needles; Neoplasms; Protocols; Computer Simulation; Connective Tissue; DNA; Drug Delivery Systems; Electrodes; Electromagnetic Fields; Electroporation; Equipment Failure Analysis; Finite Element Analysis; Gene Therapy; Models, Biological; Needles; Radiometry; Transfection;