DocumentCode :
806802
Title :
Timing System for Ge(Li)-Detectors
Author :
Akimov, Yu.K. ; Andert, K. ; Kalinin, I. ; Ortlepp, H.-G.
Author_Institution :
Joint Institute for Nuclear Research Dubna, USSR
Volume :
19
Issue :
3
fYear :
1972
fDate :
6/1/1972 12:00:00 AM
Firstpage :
404
Lastpage :
410
Abstract :
A timing system for coincidence and nuclear lifetime measurements with Ge(Li)-detectors is described. The timing discriminator, based on the ARC-principle, is designed using integrated circuits. Uncompensated pulses from low energy events near the threshold of the timing discriminator are eliminated by a special internal coincidence circuit. Output pulses caused by noise, are excluded by duration discrimination. The detector signals are amplified by a preamplifier and a fast amplifier with 2.5 and 3.5 nsec risetimes,respectively. The time resolution has been obtained 15.6 , 2.7 and 1.1 nsec FWHM for energies 50, 511 and 1330 keV, respectively, by using of a planar detector (1.9 cm3) with a charge collection time of 40 nsec. The differential peak shift in a range from 10 to 1330 keV was 1.2 nsec.
Keywords :
Detectors; Energy resolution; Integrated circuit noise; Lifetime estimation; Preamplifiers; Pulse amplifiers; Pulse circuits; Signal detection; Signal resolution; Timing;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1972.4326757
Filename :
4326757
Link To Document :