Abstract :
A circuit has been developed for an Isotope Calibrator which will accurately measure ionization chamber currents in the order of 10-14 to 10-7amps. The circuit incorporates a self-zeroing feedback network which virtually eliminates drift, as well as high impedance MOS FET switches and a MOS insulated gate FET source follower in the input circuit to minimize errors due to leakage currents.
Keywords :
Current measurement; FETs; Feedback circuits; Impedance; Insulation; Ionization chambers; Isotopes; Leakage current; Switches; Switching circuits;