• DocumentCode
    808382
  • Title

    A Computer Study of Time Resolutions and Charge Collection Times of Ge Gamma-Ray Detectors with Currently Employed Time Pick-Off Techniques

  • Author

    Bohm, C. ; Cho, Z.H.

  • Author_Institution
    Institute of Physics University of Stockholm Sweden
  • Volume
    20
  • Issue
    1
  • fYear
    1973
  • Firstpage
    246
  • Lastpage
    255
  • Abstract
    An attempt to quantitatively calculate the time resolutions from different combinations of Ge gamma-ray detectors, filters, and time pick-off techniques, is described. The study is based on charge collection time calculations previously studied, optimizations of filters, and the choice of parameters in the different time pick-off techniques. Results of the time resolution simulations with a 3 cc planar detector and a 5.6 cc coaxial detector of the Ge (Li) type, are presented and are in good agreement with previously obtained experimental and theoretical results. A brief discussion of the most common trigger modes (and related time jitters and time walks) particularly in reference to the Ge gamma-ray detectors, is also given.
  • Keywords
    Coaxial components; Energy resolution; Filters; Gamma ray detection; Gamma ray detectors; Geometry; Physics; Pulse shaping methods; Shape; Timing;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1973.4326916
  • Filename
    4326916