• DocumentCode
    808569
  • Title

    A new design method for completely open architecture permanent magnet for MRI

  • Author

    Yao, Yingying ; Fang, Youtong ; Koh, Chang Seop ; Ni, Guangzheng

  • Author_Institution
    Coll. of Electr. Eng., Zhejiang Univ., China
  • Volume
    41
  • Issue
    5
  • fYear
    2005
  • fDate
    5/1/2005 12:00:00 AM
  • Firstpage
    1504
  • Lastpage
    1507
  • Abstract
    A completely open magnet geometry satisfies the requirement of a MRI system for the surgical procedure. This paper presents a new flexible design method to facilitate the construction of a completely open MRI magnet by combining the magnetic dipole pair model and optimization algorithm. A numerical implementation shows that the method provides a main magnetic field with sufficient homogeneity and strength at the imaging region, which is defined outside of the magnet structure, and minimizes the cost of the MRI equipment.
  • Keywords
    magnetic fields; magnetic moments; magnetic resonance imaging; optimisation; permanent magnets; MRI system; completely open MRI magnet; magnet geometry; magnetic dipole pair model; magnetic field; magnetic resonance imaging; optimization algorithm; permanent magnet; Computational geometry; Computer architecture; Design methodology; Educational institutions; Magnetic devices; Magnetic fields; Magnetic resonance imaging; Permanent magnets; Superconducting magnets; Surgery; Completely open MRI magnet; magnetic dipole; optimization; permanent magnet;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/TMAG.2005.845080
  • Filename
    1430895