Title :
A new design method for completely open architecture permanent magnet for MRI
Author :
Yao, Yingying ; Fang, Youtong ; Koh, Chang Seop ; Ni, Guangzheng
Author_Institution :
Coll. of Electr. Eng., Zhejiang Univ., China
fDate :
5/1/2005 12:00:00 AM
Abstract :
A completely open magnet geometry satisfies the requirement of a MRI system for the surgical procedure. This paper presents a new flexible design method to facilitate the construction of a completely open MRI magnet by combining the magnetic dipole pair model and optimization algorithm. A numerical implementation shows that the method provides a main magnetic field with sufficient homogeneity and strength at the imaging region, which is defined outside of the magnet structure, and minimizes the cost of the MRI equipment.
Keywords :
magnetic fields; magnetic moments; magnetic resonance imaging; optimisation; permanent magnets; MRI system; completely open MRI magnet; magnet geometry; magnetic dipole pair model; magnetic field; magnetic resonance imaging; optimization algorithm; permanent magnet; Computational geometry; Computer architecture; Design methodology; Educational institutions; Magnetic devices; Magnetic fields; Magnetic resonance imaging; Permanent magnets; Superconducting magnets; Surgery; Completely open MRI magnet; magnetic dipole; optimization; permanent magnet;
Journal_Title :
Magnetics, IEEE Transactions on
DOI :
10.1109/TMAG.2005.845080