A new design method for completely open architecture permanent magnet for MRI
Author
Yao, Yingying ; Fang, Youtong ; Koh, Chang Seop ; Ni, Guangzheng
Author_Institution
Coll. of Electr. Eng., Zhejiang Univ., China
Volume
41
Issue
5
fYear
2005
fDate
5/1/2005 12:00:00 AM
Firstpage
1504
Lastpage
1507
Abstract
A completely open magnet geometry satisfies the requirement of a MRI system for the surgical procedure. This paper presents a new flexible design method to facilitate the construction of a completely open MRI magnet by combining the magnetic dipole pair model and optimization algorithm. A numerical implementation shows that the method provides a main magnetic field with sufficient homogeneity and strength at the imaging region, which is defined outside of the magnet structure, and minimizes the cost of the MRI equipment.
Keywords
magnetic fields; magnetic moments; magnetic resonance imaging; optimisation; permanent magnets; MRI system; completely open MRI magnet; magnet geometry; magnetic dipole pair model; magnetic field; magnetic resonance imaging; optimization algorithm; permanent magnet; Computational geometry; Computer architecture; Design methodology; Educational institutions; Magnetic devices; Magnetic fields; Magnetic resonance imaging; Permanent magnets; Superconducting magnets; Surgery; Completely open MRI magnet; magnetic dipole; optimization; permanent magnet;