DocumentCode :
808579
Title :
A simple method to incorporate the effects of an RF shield into RF resonator analysis for MRI applications
Author :
Jin, Jianming ; Magin, Richard L. ; Shen, Gary ; Perkins, Thomas
Author_Institution :
Dept. of Electr. & Comput. Eng., Illinois Univ., Urbana, IL, USA
Volume :
42
Issue :
8
fYear :
1995
Firstpage :
840
Lastpage :
843
Abstract :
A simple but effective method is proposed to incorporate the effects of a radiofrequency (RF) shield into the analysis of RF resonators for magnetic resonance imaging (MRI) applications. It Is shown that the method can predict the resonant frequencies of RF resonators within 5% of the measured values.
Keywords :
biomedical NMR; electromagnetic shielding; resonators; MRI applications; RF resonators analysis; RF shield effects; magnetic resonance imaging; medical diagnostic imaging; resonant frequencies prediction; Capacitors; Coils; Conductors; Equivalent circuits; Frequency measurement; Image analysis; Magnetic analysis; Magnetic resonance imaging; Radio frequency; Resonant frequency; Magnetic Resonance Imaging; Radiation Protection; Radio Waves;