DocumentCode :
808650
Title :
Performance of a Pressurized Xenon-Filled Muitiwire Proportional Chamber
Author :
Kaufman, L. ; Perez-Mendez, V. ; Stoker, G.
Author_Institution :
Lawrence Berkeley Laboratory Berkeley, California
Volume :
20
Issue :
1
fYear :
1973
Firstpage :
426
Lastpage :
428
Abstract :
We describe the performance of a multiwire proportional chamber filled with Xenon (93% Xe, 7% C02) to a pressure of 60 psi absolute. The readout is by the electromagnetic delay line method. Measurements taken at 60 keV and 140 keV show that 1 mm and 2 mm lead grid patterns can be resolved. MTF curves are given for both these energies and the projected performance for clinical work with 99Tc is described.
Keywords :
Anodes; Bars; Gamma rays; High-resolution imaging; Imaging phantoms; Laboratories; Optical imaging; Tungsten; Wires; Xenon;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1973.4326940
Filename :
4326940
Link To Document :
بازگشت