DocumentCode
808812
Title
CdTe Gamma Spectrometers for Nondestructive Analysis of Nuclear Fuels
Author
Higinbotham, W. ; Zanio, K. ; Akutagawa, W.
Author_Institution
Brookhaven National Laboratories, Upton, L. I., NY 11973
Volume
20
Issue
1
fYear
1973
Firstpage
510
Lastpage
513
Abstract
Gamma-ray spectrometry is widely used for identification and assay of nuclear fuel and scrap materials. CdTe detectors are especially promising for use by nuclear safeguards inspectors because of their small physical size and good energy resolution. A 20 mm3 CdTe detector has been used to determine the isotopic composition of uranium and plutonium samples. The energy resolution is superior to that of NaI so that the major lines of Pu-239 are more easily distinguished from interfering lines of Pu-241 and Am-241.
Keywords
Energy resolution; Gamma ray detection; Gamma ray detectors; Germanium; Laboratories; Materials testing; Nuclear fuels; Semiconductor materials; Spectroscopy; Temperature;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1973.4326955
Filename
4326955
Link To Document