DocumentCode :
809704
Title :
Comparison of the electrical stimulation performance of four cochlear implant electrodes
Author :
Choi, Charles T M ; Lai, Wei-Dian ; Chen, Yu-Bin
Author_Institution :
Electr. Eng. Dept., I-Shou Univ., Kaohsiung, Taiwan
Volume :
41
Issue :
5
fYear :
2005
fDate :
5/1/2005 12:00:00 AM
Firstpage :
1920
Lastpage :
1923
Abstract :
This paper proposes a method to compare the neural excitation of four different cochlear implant electrodes by coupling finite-element analysis and computational neuroscience method. The goal is to identify the type of electrodes and electrode positions that will maximize the neural excitation of the human cochlea. Electrodes are moved translationally and rotationally in the inner cochlea based on straight and circular cochlea models. The four electrodes modeled are the banded electrodes, half banded electrodes, planar electrodes, and ball electrodes, and they were rotated from +180° to -180°. The planar electrode generates the highest stimulation of the four electrodes for a given electrical power input. This paper also proposes means to compute directions which the electrode should point to in order to maximize the neural excitation of a human cochlea.
Keywords :
biomedical electrodes; finite element analysis; neuromuscular stimulation; prosthetics; activating function; ball electrodes; cochlea models; cochlear implant electrodes; computational neuroscience method; electrical stimulation performance; finite-element analysis; half banded electrodes; human cochlea; neural excitation; planar electrodes; Biomembranes; Cochlear implants; Electrical stimulation; Electrodes; Finite element methods; Humans; Nerve fibers; Neuroscience; Power generation; Spirals; Activating function; cochlear implant electrodes; electrical stimulation; finite-element analysis;
fLanguage :
English
Journal_Title :
Magnetics, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9464
Type :
jour
DOI :
10.1109/TMAG.2005.846228