DocumentCode
813589
Title
Phase digitizing sharpens timing measurements
Author
Chu, David
Author_Institution
Hewlett-Packard Co., Santa Clara, CA, USA
Volume
25
Issue
7
fYear
1988
fDate
7/1/1988 12:00:00 AM
Firstpage
28
Lastpage
32
Abstract
Oscilloscopes and other instruments that so precisely measure voltage are ill-suited to handling time-based communications signals, for their preoccupation with amplitude compromises timing accuracy, wastes memory, and complicates timing analysis. A measurement technique is presented, called phase digitizing, that remedies the situation by continuously counting a signal´s zero-crossings and recording exactly when they occur. The deficiencies of previous approaches are examined for the case of spread-spectrum measurements. The principles and characteristics of the approach are described.<>
Keywords
digital instrumentation; phase measurement; spread spectrum communication; phase digitizing; signal zero-crossings; spread-spectrum measurements; timing measurements; voltage measurement; Accuracy; Instruments; Measurement techniques; Oscilloscopes; Phase measurement; Signal analysis; Spread spectrum communication; Timing; Voltage measurement; Waste handling;
fLanguage
English
Journal_Title
Spectrum, IEEE
Publisher
ieee
ISSN
0018-9235
Type
jour
DOI
10.1109/6.4572