• DocumentCode
    813904
  • Title

    Low Temperature Characteristics of Ion Implanted Silicon Position Sensitive Detectors

  • Author

    Elad, E. ; Sareen, R.

  • Author_Institution
    ORTEC, Inc. 100 Midland Road Oak Ridge, Tennessee 37830
  • Volume
    21
  • Issue
    1
  • fYear
    1974
  • Firstpage
    75
  • Lastpage
    84
  • Abstract
    The operation of ion implanted silicon position sensitive detectors at low temperatures will be described. The improved noise performance of the detector at these temperatures in both the energy and position modes will be analyzed and its implications on low energy applications will be discussed. A large part of the investigation was devoted to the study of boron implanted resistors and their noise properties both at room temperature and cryogenic temperatures.
  • Keywords
    Boron; Energy resolution; Linearity; Performance analysis; Position sensitive particle detectors; Radiation detectors; Resistors; Semiconductor device noise; Silicon; Temperature sensors;