• DocumentCode
    814760
  • Title

    Delay Line Readouts for High Purity Germanium Medical Imaging Cameras

  • Author

    Kaufman, Leon ; Camp, David C. ; McQuaid, James H. ; Armantrout, Guy A. ; Swierkowski, Steve P. ; Lee, Kai

  • Author_Institution
    University of California, San Francisco, Ca
  • Volume
    21
  • Issue
    1
  • fYear
    1974
  • Firstpage
    652
  • Lastpage
    657
  • Abstract
    High purity germanium offers excellent potential for use in nuclear medical imaging cameras. A position and energy readout technique using two inexpensive delay lines has been developed for these cameras. Results obtained with a 1-cm2 , 4 mm deep, 5×5 strip high purity germanium detector are 2.1 mm full width spatial resolution, a measured single strip resolution of 0.65 mm full width half maximum (FWHM), a 25 element uncollimated energy resolution of 2.95 keV FWHM, and 2.65 keV FWHM for a single central element at 140 keV.
  • Keywords
    Biomedical imaging; Cameras; Delay lines; Detectors; Energy resolution; Face detection; Germanium; Spatial resolution; Strips; Wires;
  • fLanguage