• DocumentCode
    814818
  • Title

    A Scintillation Camera

  • Author

    Blum, Alvin S.

  • Author_Institution
    Broward General Medical Center Fort Lauderdale, Florida 33316
  • Volume
    21
  • Issue
    1
  • fYear
    1974
  • Firstpage
    695
  • Lastpage
    696
  • Abstract
    A stationary radionuclide imaging system employs a planar scintillation crystal viewed by multiple photodetectors. A new positioning technique allows use of signal information that must be discarded by conventional cameras. A calibrator generates photodetector signal patterns characteristic of known radiation source locations. Unknown sources are positioned by matching signal patterns of their radiations with the stored calibration signal patterns. The integral of individual location is displayed as a corrected image.
  • Keywords
    Calibration; Cameras; Data processing; Libraries; Optical scattering; Photodetectors; Photonic crystals; Position measurement; Signal generators; Signal processing;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1974.4327534
  • Filename
    4327534