DocumentCode :
814837
Title :
Chronic in Vivo Testing of a Fully Implantable Intracranial Pressure Sensor
Author :
Meyer, Glenn A. ; Lyon, Warren C. ; Bustard, Thomas S.
Author_Institution :
Department of Neurosurgery Medical College of Wisconsin 8700 West Wisconsin Avenue Milwaukee, Wisconsin
Volume :
21
Issue :
1
fYear :
1974
Firstpage :
702
Lastpage :
706
Abstract :
This paper describes the authors´ approach to the problem of long term intracranial pressure measurement. Several prior devices have not proven useful for clinical studies of more than a few days duration. Study of 22 of the authors´ devices in primates has established the validity of both engineering and medical design assumptions. Useful pressure data has been collected for up to 2 months. Redesign to decrease tambour permeability should allow a useful life of months or years.
Keywords :
Batteries; Capacitors; Cranial pressure; In vivo; Intracranial pressure sensors; Scalp; Switches; Temperature; Testing; Transducers;