DocumentCode
814956
Title
Optimization of the Clipping Time Constant for BF3 Neutron Detectors Operated in a High Level Gamma Environment
Author
Brown, Donald P.
Author_Institution
Hanford Engineering Development Laboratory Westinghouse Hanford Company Richland, Washington
Volume
21
Issue
1
fYear
1974
Firstpage
763
Lastpage
766
Abstract
Radiation-hardened BF3 neutron detectors (proportional counters) have been selected for use as delayed neutron detectors in the failed fuel monitoring system of the Fast Flux Test Facility (FFTF) being constructed by the Westinghouse Hanford Company at Richland, Washington. These detectors must exhibit fpproximately 6 to 7 counts/sec (cps) per neutron/cm2-sec (nv) for neutrons while operating in a gamma environment of a about 1000 R/hr; thus, electronic counting systems capable of providing the best possible sensitivity must be used. In addition to exhibiting appropriate charge gain and sufficiently low noise, the preamplifier (or amplifier) must include at least one stage of differentiation to reduce gamma interference (pulse pileup) effects. The selection of the time constant depends on the rise time of the neutron-caused pulse from the BF3 tube and on the maximum gamma level in which the tube must operate. The conflicting requirements of large time constant to maintain pulse amplitude and small time constant to reduce gamma effects lead to an optimum value to achieve the best possible neutron detection sensitivity.
Keywords
Condition monitoring; Delay; Fuels; Gamma ray detection; Gamma ray detectors; Neutrons; Pulse amplifiers; Radiation detectors; Radiation monitoring; Test facilities;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1974.4327546
Filename
4327546
Link To Document