Title :
A Stable ADC for Use with Ge(Li) Detectors
Author_Institution :
Jet Propulsion Laboratory 4800 Oak Grove Drive Pasadena, CA 91103 (213) 354-6174
Abstract :
A new type of ADC is being developed to satisfy the requirements of the HEAO high spectral resolution gamma ray spectrometer. This 13 bit design utilizes a double successive approximation technique that overcomes the usual differential linearity problem. Stability of a few ppm/°C, differential linearity of 2% and integral linearity of 0.05% have been demonstrated.
Keywords :
Detectors; Laboratories; Linearity; Propulsion; Pulse amplifiers; Pulse shaping methods; Resistors; Spectroscopy; Stability; Voltage;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :