Title :
Fabrication and X-Ray Calibration of Thin Plastic Scintillator Detectors
Author :
Lyons, P.B. ; Lier, D.W.
Author_Institution :
University of California Los Alamos Scientific Laboratory Los Alamos, New Mexico 87544
Volume :
22
Issue :
1
fYear :
1975
Firstpage :
88
Lastpage :
92
Abstract :
Thin film plastic scintillator detectors were fabricated and calibrated with x rays in the 1.5-20 keV range. The calibration systems used are described. Measurements showed that the presence of a film substrate can significantly change the detector response. X-ray conversion efficiency was determined as a function of energy and shown to be a strong function of energy for low energy x rays.
Keywords :
Anodes; Calibration; Fabrication; Filters; Fluorescence; Plastics; Vacuum systems; Voltage; X-ray detection; X-ray detectors;