DocumentCode :
815865
Title :
Comparitive Ionization Energies for Protons, Deuterons and Alpha Particles in High Purity Germanium and Si(Li) Nuclear Radiation Detectors
Author :
Martini, M. ; Raudorf, T.W. ; Stott, W.R. ; Waddington, J.C.
Author_Institution :
Nuclear Radiation Developments, Division of Electronic Associates of Canada Limited
Volume :
22
Issue :
1
fYear :
1975
Firstpage :
145
Lastpage :
148
Abstract :
Relative measurements of the average energy, ¿, expended for electron-hole pair generation in silicon and high purity germanium detectors were made using 10 to 19 MeV proton, deuteron and alpha particle beams. The results show that ¿ is constant for different particles in high purity germanium detectors but not in silicon detectors. A model is presented which could explain both our results and those published in the literature.
Keywords :
Alpha particles; Energy measurement; Germanium; Ionization; Nuclear power generation; Particle beam measurements; Particle measurements; Protons; Radiation detectors; Silicon;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1975.4327632