DocumentCode :
816052
Title :
X-Y Multiwire Proportional Chamber (MWPC) with Delay Line Readout for the Magnet-Spectrograph of the "Julic" Cyclotron
Author :
Kohler, Mark ; Muller, K.D. ; Teske, M.
Author_Institution :
Zentrallabor fÿr Elektronik KFA Jÿlich, D 517 Jÿlich, Germany
Volume :
22
Issue :
1
fYear :
1975
Firstpage :
272
Lastpage :
276
Abstract :
A MWPC with an active area of 30 × 4 cm2 has been developed using two-dimensional delay-line readout. To avoid the problems associated with long wires, the anode and cathode wires are oriented at 45° to the X and Y axes; this is important for further extension in one direction. New electromagnetic delay lines for wire chamber readout with high bandwidth and low attenuation are used. The X-Y spatial resolution has been measured to be about 0,5 mm FWHM.
Keywords :
Anodes; Attenuation; Bandwidth; Cathodes; Cyclotrons; Delay lines; Detectors; Kinematics; Spatial resolution; Wires;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1975.4327650