Title
Tomographic Reconstruction from Fan Beam Geometry Using Radon´s Integration Method
Author
Beattie, John W.
Author_Institution
Memorial Sloan-Kettering Cancer Center New York, New York 10021
Volume
22
Issue
1
fYear
1975
Firstpage
359
Lastpage
363
Abstract
Tomographic scanners of the parallel beam and fan beam type have differences in the direction of sweep, r vs. ¿ respectively. This leads to different options in the "fill" of the sampling space, which may be important in accurate reconstruction. These differences are denoted in reference to the Radon integrational method of reconstruction. An algorithm using source angle gating of data from a fan beam scanner is given with a reconstruction from a simulation phantom scan which used discrete source and detector apertures with rotation smearing.
Keywords
Apertures; Cancer; Computational modeling; Envelope detectors; Geometry; Imaging phantoms; Sampling methods; Strips; Surface reconstruction; Tomography;
fLanguage
English