• DocumentCode
    816677
  • Title

    A Closed Loop Threshold Calibrator for Pulse Height Discriminators

  • Author

    Peletier, Daniel P.

  • Author_Institution
    The Johns Hopkins University Applied Physics Laboratory Silver Spring, Maryland
  • Volume
    22
  • Issue
    1
  • fYear
    1975
  • Firstpage
    592
  • Lastpage
    594
  • Abstract
    This paper describes a simple method of inflight calibration of pulse amplitude discriminator thresholds on satellite-borne particle detection experiments. The calibrator uses the discriminator output as a feedback control signal to adjust the average output of a test pulse generator. The test pulse amplitude is stabilized when the discriminator "half-triggers". The average control voltage is a direct measure of the pulse height discriminator threshold. Better than 1% accuracy is obtained over a 50:1 range of discriminator settings. Automatic ranging circuits extend the calibrator dynamic range to 5000:1.
  • Keywords
    Calibration; Choppers; Circuit testing; Clocks; Detectors; Instruments; Preamplifiers; Pulse amplifiers; Pulse measurements; Threshold voltage;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1975.4327709
  • Filename
    4327709