DocumentCode :
818920
Title :
High Resolution Beam Profile Measurements Using Phosphorescence
Author :
Bjarngard, B.E. ; Chen, G.T.Y. ; Maddox, B.J.
Author_Institution :
Joint Center for Radiation Therapy and Dept. of Radiation Therapy Harvard Medical School Boston, Mass. 02115
Volume :
22
Issue :
3
fYear :
1975
fDate :
6/1/1975 12:00:00 AM
Firstpage :
1558
Lastpage :
1560
Abstract :
Commercially available thermoluminescent material, CaSO4: Mn incorporated in Teflon tape, is used in a phosphoresecence decay mode to measure beam profiles. The light image resulting from radiation exposure is scanned by a microscope-photomultiplier system having a spatial resolution of 0.25 mm. Several applications to various types of accelerator beams are presented.
Keywords :
Apertures; Biomedical applications of radiation; Glass; Microscopy; Optical fibers; Particle beams; Phosphorescence; Phosphors; Photomultipliers; Temperature;