DocumentCode
819106
Title
Online dosimetry based on optically stimulated luminescence materials
Author
Vaillé, J.R. ; Ravotti, F. ; Garcia, P. ; Glaser, M. ; Matias, S. ; Idri, K. ; Boch, J. ; Lorfèvre, E. ; McNulty, P.J. ; Saigné, F. ; Dusseau, L.
Author_Institution
TS & PH Dept., CERN, Geneva, Switzerland
Volume
52
Issue
6
fYear
2005
Firstpage
2578
Lastpage
2582
Abstract
A version of the Optically Stimulated Luminescence (OSL) sensor specifically developed to monitor the dose online in radiation facilities is presented and calibrated with 60Co. The lowest dose measurable at the extremity of a 20 m cable is 0.3 mGy.
Keywords
calibration; dosimetry; photodetectors; radiation monitoring; 0.3 mGy; 20 m; 60Co; calibration; online dose monitoring; online dosimetry; optically stimulated luminescence materials; optically stimulated luminescence sensor; radiation facilities; Dosimetry; Extraterrestrial measurements; Large Hadron Collider; Luminescence; Optical materials; Optical saturation; Optical sensors; Protons; Radiation monitoring; Stimulated emission; Online dosimetry; Optically Stimulated Luminescence (OSL); protons;
fLanguage
English