• DocumentCode
    819106
  • Title

    Online dosimetry based on optically stimulated luminescence materials

  • Author

    Vaillé, J.R. ; Ravotti, F. ; Garcia, P. ; Glaser, M. ; Matias, S. ; Idri, K. ; Boch, J. ; Lorfèvre, E. ; McNulty, P.J. ; Saigné, F. ; Dusseau, L.

  • Author_Institution
    TS & PH Dept., CERN, Geneva, Switzerland
  • Volume
    52
  • Issue
    6
  • fYear
    2005
  • Firstpage
    2578
  • Lastpage
    2582
  • Abstract
    A version of the Optically Stimulated Luminescence (OSL) sensor specifically developed to monitor the dose online in radiation facilities is presented and calibrated with 60Co. The lowest dose measurable at the extremity of a 20 m cable is 0.3 mGy.
  • Keywords
    calibration; dosimetry; photodetectors; radiation monitoring; 0.3 mGy; 20 m; 60Co; calibration; online dose monitoring; online dosimetry; optically stimulated luminescence materials; optically stimulated luminescence sensor; radiation facilities; Dosimetry; Extraterrestrial measurements; Large Hadron Collider; Luminescence; Optical materials; Optical saturation; Optical sensors; Protons; Radiation monitoring; Stimulated emission; Online dosimetry; Optically Stimulated Luminescence (OSL); protons;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2005.860707
  • Filename