Characterization and Simulation of a New Design Parallel-Plate Ionization Chamber for CT Dosimetry at Calibration Laboratories
Author :
Perini, Ana P. ; Neves, Lucio P. ; Maia, Ana F. ; Caldas, L.V.E.
Author_Institution :
Inst. de Pesquisas Energeticas e Nucl., Comissao Nac. de Energia Nucl. (IPEN-CNEN/SP), Sao Paulo, Brazil
Volume :
60
Issue :
6
fYear :
2013
fDate :
Dec. 2013
Firstpage :
4641
Lastpage :
4649
Abstract :
In this work, a new extended-length parallel-plate ionization chamber was tested in the standard radiation qualities for computed tomography established according to the half-value layers defined at the IEC 61267 standard, at the Calibration Laboratory of the Instituto de Pesquisas Energéticas e Nucleares (IPEN). The experimental characterization was made following the IEC 61674 standard recommendations. The experimental results obtained with the ionization chamber studied in this work were compared to those obtained with a commercial pencil ionization chamber, showing a good agreement. With the use of the PENELOPE Monte Carlo code, simulations were undertaken to evaluate the influence of the cables, insulator, PMMA body, collecting electrode, guard ring, screws, as well as different materials and geometrical arrangements, on the energy deposited on the ionization chamber sensitive volume. The maximum influence observed was 13.3% for the collecting electrode, and regarding the use of different materials and design, the substitutions showed that the original project presented the most suitable configuration. The experimental and simulated results obtained in this work show that this ionization chamber has appropriate characteristics to be used at calibration laboratories, for dosimetry in standard computed tomography and diagnostic radiology quality beams.
Keywords :
Monte Carlo methods; calibration; dosimetry; electrodes; insulators; ionisation chambers; CT dosimetry; IEC 61267 standard; IEC 61674 standard recommendations; IPEN; Instituto de Pesquisas Energeticas e Nucleares; PENELOPE Monte Carlo code; PMMA body; cable influence; calibration laboratories; commercial pencil ionization chamber; deposited energy; design use; diagnostic radiology quality beams; electrode influence; experimental characterization; extended-length parallel-plate ionization chamber; geometrical arrangement; guard ring; half-value layers; insulator influence; ionization chamber sensitive volume; material arrangement; material use; maximum observed influence; parallel-plate ionization chamber design characterization; parallel-plate ionization chamber design simulation; screw influence; standard computed tomography; standard radiation qualities; Computed tomography; Dosimetry; IEC standards; Ionization chambers; Monte Carlo methods; Semiconductor process modeling; Simulation; Computed tomography; Monte Carlo simulation; dosimetry; operational tests; parallel-plate ionization chamber;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2013.2283798
Filename :
6655998
Link To Document :
بازگشت