DocumentCode
820203
Title
A Pulse Processor for X-Ray Spectrometry with Si(Li)-Detectors
Author
Kandiah, K. ; Smith, A.J. ; White, G.
Author_Institution
United Kingdom Atomic Energy Authority Harwell, Didcot, Oxon, U.K.
Volume
22
Issue
5
fYear
1975
Firstpage
2058
Lastpage
2065
Abstract
This paper describes the general features of a pulse processor for use with opto-electronic charge sensitive head amplifiers particularly with Si(Li) X-ray detectors. The instrument contains time-variant filters with direct coupled amplifiers and comprehensive pile-up protection and live time correction. The test results indicate that the system can cope with extremely wide range of signal processing times, counting rates and changes of spectral content without any appreciable degradation in energy resolution and with useful output rates for input rates up to well over 50,000 events/sec.
Keywords
Degradation; Energy resolution; Filters; Instruments; Protection; Pulse amplifiers; Signal processing; Spectroscopy; System testing; X-ray detectors;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1975.4328065
Filename
4328065
Link To Document