• DocumentCode
    820203
  • Title

    A Pulse Processor for X-Ray Spectrometry with Si(Li)-Detectors

  • Author

    Kandiah, K. ; Smith, A.J. ; White, G.

  • Author_Institution
    United Kingdom Atomic Energy Authority Harwell, Didcot, Oxon, U.K.
  • Volume
    22
  • Issue
    5
  • fYear
    1975
  • Firstpage
    2058
  • Lastpage
    2065
  • Abstract
    This paper describes the general features of a pulse processor for use with opto-electronic charge sensitive head amplifiers particularly with Si(Li) X-ray detectors. The instrument contains time-variant filters with direct coupled amplifiers and comprehensive pile-up protection and live time correction. The test results indicate that the system can cope with extremely wide range of signal processing times, counting rates and changes of spectral content without any appreciable degradation in energy resolution and with useful output rates for input rates up to well over 50,000 events/sec.
  • Keywords
    Degradation; Energy resolution; Filters; Instruments; Protection; Pulse amplifiers; Signal processing; Spectroscopy; System testing; X-ray detectors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1975.4328065
  • Filename
    4328065