DocumentCode
822616
Title
Measurement of Absolute Lung Density by Compton-Scatter Densitometry
Author
Kaufman, Leon ; Gamsu, Gordon ; Savoca, Charles ; Swann, Sybil ; Murphey, Louis ; Hruska, Bernard ; Palmer, David ; Ullman, John
Author_Institution
University of California, San Francisco
Volume
23
Issue
1
fYear
1976
Firstpage
599
Lastpage
605
Abstract
The non-invasive determination of absolute lung density would be of great importance if it could be used in the evaluation of patients with pulmonary edema. To obtain this measurement, we have developed a a portable densitometer using CdTe room-temperature detectors, compact collimators made of tantalum tubes and a Gd-153 source. System response in the density range of 0.1 to 1 g/cm3 is linear. Variations in the absorption paths of up to 8 cm of lucite do not produce significant changes on measured density values.
Keywords
Attenuation; Collimators; Density measurement; Detectors; Electromagnetic scattering; Electromagnetic wave absorption; Lungs; Niobium; Organic materials; Particle scattering;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1976.4328313
Filename
4328313
Link To Document