DocumentCode
822836
Title
Proportional Counter for Monitoring Low Levels of Krypton-85 and Other Beta- and X-Ray-Emitting Nuclides
Author
Gregory, J.C. ; Parnell, T.A.
Author_Institution
The University of Alabama in Huntsville, Huntsville, AL 35807
Volume
23
Issue
1
fYear
1976
Firstpage
715
Lastpage
718
Abstract
The use of the inert radioactive gas Kr-85 for tracer studies, leak detection, etc., will become more widespread as the sensitivity of available detectors increases. A multiwire multiplane proportional counter system for beta detection has been developed with a measured minimum detectable concentration for a 3-minute count of 3 Ã 10-3 pCi cm-3 of Kr-85 in a semi-infinite cloud. Larger proportional counters operated for counting times of several hours can reduce this limit to below 1 Ã 10-4 pCi cm-3. These detectors should be useful in monitoring low levels of -the radioactive inert gases and of x-ray emitting nuclides.
Keywords
Calibration; Counting circuits; Gases; Laboratories; Leak detection; Monitoring; NASA; Radiation detectors; X-ray detection; X-ray detectors;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1976.4328333
Filename
4328333