DocumentCode :
822871
Title :
Evaluation of Standardized Ge(Li) Gamma Ray Detectors for Low Level Environmental Measurements
Author :
Walford, G.V. ; Cooper, J.A. ; Keyser, R.M.
Author_Institution :
ORIEC, Incorporated 100 Midland Road Oak Ridge, Tenessee 37830
Volume :
23
Issue :
1
fYear :
1976
Firstpage :
734
Lastpage :
742
Abstract :
Counting sensitivity for stardardized Ge(Li) detectors has been determined for a large number of detector-sample geometries and counting conditions. It has been found that relatively simple systems can satisfy a majority of applications, thereby resulting in a favorable cost-effective approach to enviromental measurements. Recommendations are made for the improvement of significant counting parameters for speific applications, resulting in the maximum counting capability for a given detector system.
Keywords :
Costs; Energy measurement; Energy resolution; Gamma ray detectors; Inductors; Laboratories; Power measurement; Power system protection; Scattering; Testing;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1976.4328336
Filename :
4328336
Link To Document :
بازگشت