Title :
Emittance Measurements on Negative Ion Sources
Author :
Doucas, By G. ; Greenway, T.J.L. ; Hyder, H.R.McK. ; Knox, A.B.
Author_Institution :
Nuclear Physics Laboratory, Oxford University, England
fDate :
4/1/1976 12:00:00 AM
Abstract :
An automated system has been developed for measuring the emittance of beams from negative ion sources. The data are stored as two-dimensional spectra in a kicksorter and subsequently analysed with a PDP-10 computer. The system was first used to measure the properties of, and assess the value of design changes to, a Middleton sputter source. Subsequently, the emittance of several different types of negative ion source have been compared. A retarding field energy analyzer of high resolution has been used to measure the beam energy spreads from the Middleton sputter source and from other negative ion sources, including a direct extraction duoplasmatron and a sputter Penning source. Typical spectra have a width of a few eV, with a small group of higher energy ions.
Keywords :
Brightness; Current measurement; Data acquisition; Energy measurement; Ion sources; Laboratories; Nuclear measurements; Nuclear physics; Particle beams; Velocity measurement;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1976.4328425