DocumentCode :
824929
Title :
2D imaging of ultrasound fields using CCD array to map output of Fabry-Perot polymer film sensor
Author :
Lamont, M. ; Beard, P.C.
Author_Institution :
Dept. of Med. Phys. & Bioeng., Univ. Coll. London, UK
Volume :
42
Issue :
3
fYear :
2006
Firstpage :
187
Lastpage :
189
Abstract :
A 2D ultrasound field imaging system that uses a CCD array to map the optical output of a planar Fabry-Perot polymer film sensing interferometer is described. This system enables ultrasound fields to be rapidly mapped over a 6.6×4.7 mm aperture with spatial and temporal resolutions of the order of 10 μm and 10 ns, respectively. The concept was successfully demonstrated by imaging the acoustic field at the focus of a pulsed 5 MHz PZT ultrasound transducer. The technique offers an alternative to piezoelectric-based methods for high-resolution biomedical and industrial ultrasonic imaging and field characterisation applications.
Keywords :
CCD image sensors; Fabry-Perot interferometers; biomedical imaging; piezoelectric transducers; polymer films; ultrasonic imaging; ultrasonic transducers; 10 micron; 10 ns; 2D ultrasound field imaging system; 5 MHz; CCD arrays; Fabry-Perot polymer film sensor; PZT ultrasound transducers; acoustic field imaging; biomedical ultrasonic imaging; high-resolution ultrasonic imaging; industrial ultrasonic imaging; planar Fabry-Perot polymer film sensing interferometer;
fLanguage :
English
Journal_Title :
Electronics Letters
Publisher :