Title :
Diamond dosimeter for x-ray and δ-radiation
Author :
Kozlov, S.F. ; Konorova, A. ; Kuznetsov, L.A. ; Salikov, Y.A. ; Redko, V.I. ; Grinberg, V.R. ; Meilman, M.L.
Author_Institution :
P.N. Lebedev Physical Institute of the USSR Academy of Sciences Moscow
Abstract :
Some characteristics of diamond detectors have been investigated with a view to finding out the possibilities of their application for X-ray and δ-radiation dosimetry in medical radiology. The results of this investigation are described. The design of a tissue-equivalent diamond dosimeter is shown. It possesses a uniform energy dependence and threshold sensitivity of lμR/s to the dose rate.
Keywords :
Annealing; Boron; Charge carriers; Current measurement; Current-voltage characteristics; Dark current; Gold; Polarization; Radiation detectors; Voltage measurement;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :