DocumentCode
826209
Title
Sub-Nanosecond Plastic Scintillator Time Response Studies Using Laser Produced X-Ray Pulsed Excitation
Author
Tirsell, K.G. ; Tripp, G.R. ; Lent, E.M. ; Lerche, R.A. ; Cheng, J.C. ; Hocker, L. ; Lyons, P.B.
Author_Institution
Lawrence Livermore Laboratory
Volume
24
Issue
1
fYear
1977
Firstpage
250
Lastpage
254
Abstract
The light emission time response of quenched NE111 plastic scintillators has been measured using a streak camera (20 ps resolution) and 100 to 180 ps, 1.06 ¿m, laser-produced, pulsed, low energy x-ray excitation. Each light output pulse was obtained by deconvolution from the film data using the x-ray temporal response measured with an x-ray sensitive streak camera (10 ps resolution). Time response parameters are presented for benzophenone and acetophenone, quenching agents which most effectively reduce the decay time of the singlet component. Full width-half-maximums ¿ 260 ps were observed for NE111 samples quenched with ¿ 2% benzophenone. Results are given for unquenched samples consisting of different concentrations of butyl-PBD in PVT and for the phosphor ZnO doped with Ga.
Keywords
Cameras; Energy measurement; Energy resolution; Laser excitation; Optical pulses; Plastics; Pulse measurements; Time factors; Time measurement; X-ray lasers;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1977.4328684
Filename
4328684
Link To Document