DocumentCode :
827101
Title :
Characterization of Platinum Self-Powered Detectors
Author :
Lynch, G.F. ; Shields, R.B. ; Coulter, P.G.
Author_Institution :
Atomic Energy of Canada Limited Chalk River Nuclear Laboratories Chalk River, Ontario, Canada
Volume :
24
Issue :
1
fYear :
1977
Firstpage :
692
Lastpage :
695
Abstract :
Because platinum self-powered flux detectors have a mixed (neutron and gamma-ray) response in reactor radiation fields, an experimental program was undertaken to characterize these devices. This paper describes experiments performed to determine the response parameters for Pt detectors in a heavy-water moderated reactor. Absolute y-ray and neutron sensitivities were derived from these measurements for detectors with three different emitter diameters, and the results are comrpared with previously reported values.
Keywords :
Delay; Event detection; Gamma ray detection; Gamma ray detectors; Gas detectors; Inductors; Neutrons; Platinum; Signal detection; Signal processing;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1977.4328769