• DocumentCode
    832396
  • Title

    Position-Sensitive Gas Proportional Chambers

  • Author

    Morris, C.L. ; Thiessen, H.A. ; Hoffmann, G.W.

  • Author_Institution
    Los Alamos Scientific Laboratory Los Alamos, New Mexico 87545
  • Volume
    25
  • Issue
    1
  • fYear
    1978
  • Firstpage
    141
  • Lastpage
    143
  • Abstract
    A number of multiwire, position-sensitive, gas proportional chambers have been constructed which use fast delay line (2.5 ns/cm) readouts to provide trajectory position and angle information for particles near the focal planes of the Energetic Pion Channel and Spectrometer (EPICS) and High Resolution Spectrometer (HRS) at the Clinton P. Anderson Meson Physics Facility (LAMPF). These chambers are two dimensional with anode wires in one direction and cathode wires in the other direction. When operated as ordinary delay line chambers, position resolution of 0.5 mm (FWHM) can be obtained in the cathode direction for short cathode wire lengths. If the drift information from the anodes is used, resolution of less than 0.35 mm (FWHM) can be obtained in this direction. The chambers can be operated at count rates of 5 × 105 pulses/sec with live times of greater than 90%. Efficiences, measured at lower counting rates, are typically greater than 99%.
  • Keywords
    Aluminum; Anodes; Attenuation; Cathodes; Delay lines; Logic; Position measurement; Voltage; Wire; Wounds;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1978.4329293
  • Filename
    4329293