Evaluation of Well-Type Ge(Li) Detectors for Low-Level Radiochemical Analysis
Author
Brauer, F.P. ; Mitzlaff, W.A.
Author_Institution
Battelle, Pacific Northwest Laboratories Richland, Washington 99352
Volume
25
Issue
1
fYear
1978
Firstpage
398
Lastpage
403
Abstract
Well-type Ge(Li) detectors have been evaluated for low-level gamma-ray spectrometry and radiochemical analyses. The detectors were found to have good resolution, high peak-to-Compton ratios and low backgrounds. The use of an anticoincidence shield further improves the detector performance. The detector efficiencies and backgrounds are compared with those obtained with other Ge(Li) detector systems. The well detectors were found to have better detection efficiencies and as low backgrounds as either large coaxial detectors or opposed detector systems. Sum-coincidence effects are more pronounced in the well detector and use of this feature is discussed. Applications which utilize the low-energy response of the detector are described. Minimum detectable activity levels were determined for several nuclides.
Keywords
Coaxial components; Energy measurement; Energy resolution; Gamma ray detection; Gamma ray detectors; Geometry; Laboratories; Loss measurement; Performance evaluation; Spectroscopy;