DocumentCode
832908
Title
Deuterated liquid scintillator (NE230) as a fast neutron detector for cold-fusion and other research
Author
Roberts, D.A. ; Becchetti, F.D. ; Ashktorab, K. ; Stewart, D. ; Jänecke, J. ; Gustafson, H.R. ; Dueweke, M.J.
Author_Institution
Dept. of Phys., Michigan Univ., Ann Arbor, MI, USA
Volume
39
Issue
4
fYear
1992
fDate
8/1/1992 12:00:00 AM
Firstpage
532
Lastpage
535
Abstract
Deuterated scintillators (NE230) can provide, without time-of-flight, usable neutron energy spectra for neutrons E n⩾1 MeV. The authors have assembled and calibrated several of these detectors and used them in experiments searching for d+d fusion-neutron emission (E n, 2.5 MeV) in Pd+D 2O electrolysis and Ti+D2 adsorption at LN temperatures. The detectors yielded direct fusion-neutron spectra and set limits in the Pd+D and Ti+D experiments of <7×10-24 fusion n/s/dd pair and <3×10-24 fusion n/s/dd pair respectively. In addition, these detectors have been used for in-beam accelerator experiments requiring coincidence between fast neutrons and scattered ions
Keywords
electrochemistry; neutron detection and measurement; nuclear fusion; scintillation counters; LN temperatures; NE230; Pd+D2O electrolysis; Ti+D2 adsorption; calibrated; cold-fusion; d+d fusion-neutron emission; deuterated liquid scintillator; direct fusion-neutron spectra; fast neutron detector; fast neutrons; in-beam accelerator experiments; neutron energy spectra; scattered ions; Assembly; Calibration; Detectors; Electrochemical processes; Energy measurement; Ion accelerators; Light scattering; Neutrons; Physics; Temperature;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.159660
Filename
159660
Link To Document